DocumentCode :
726239
Title :
Traceable phase calibration of a wide-bandwidth microwave Vector Signal Analyzer
Author :
Vanden Bossche, Marc ; Verbeyst, Frans ; Samant, Abhay
Author_Institution :
Network Anal. Center of Excellence, Nat. Instrum., Zaventem, Belgium
fYear :
2015
fDate :
17-22 May 2015
Firstpage :
1
Lastpage :
4
Abstract :
Numerous wide-bandwidth and multi-channel applications benefit from accurate test and measurement equipment which minimize the distortion of the generated and analyzed signals. Different techniques are described in literature to achieve this for Vector Signal Analyzers. This paper presents a simple traceable phase calibration technique for wide-bandwidth Vector Signal Analyzers using a comb generator which is able to realize a dense frequency grid up to 40 GHz and above. The comb generator itself is calibrated using a technique which is traceable to electro-optical sampling as primary standard. The calibration procedure is explained in detail showing the characterized phase distortion for different center frequencies and different IF bandwidths of a commercial VSA.
Keywords :
calibration; electro-optical effects; microwave measurement; phase measurement; VSA; comb generator; electrooptical sampling; phase distortion; traceable phase calibration technique; wide-bandwidth microwave vector signal analyzer; Extraterrestrial measurements; Frequency measurement; Phase modulation; Radio frequency; Reflection; Calibration; RF signals; comb generator; modulation; phase distortion; vector signal analyzer;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Microwave Symposium (IMS), 2015 IEEE MTT-S International
Conference_Location :
Phoenix, AZ
Type :
conf
DOI :
10.1109/MWSYM.2015.7167125
Filename :
7167125
Link To Document :
بازگشت