Title :
A wireless implantable drug infusion system with integrated dosing sensors
Author :
Sheybani, R. ; Meng, E.
Author_Institution :
Biomed. Microsyst. Lab., Univ. of Southern California, Los Angeles, CA, USA
Abstract :
A wireless electrolysis-based drug delivery system with integrated electrochemical dosing sensors suitable for management of chronic conditions is presented. Repeatable delivery performance for identical and variable bolus volumes and flow rates of phosphate buffer saline (PBS) was demonstrated. Successful delivery and sensing of 17.75 ± 0.48 μL boluses (1.78% of reservoir fill volume) of lidocaine was achieved.
Keywords :
biomedical equipment; drug delivery systems; electrochemical sensors; electrolysis; organic compounds; bolus volumes; chronic condition management; integrated electrochemical dosing sensors; lidocaine; phosphate buffer saline flow rates; wireless electrolysis-based drug delivery system; wireless implantable drug infusion system; Drug delivery; Drugs; Micropumps; Reservoirs; Sensors; Wireless sensor networks; Wires; Electrochemical sensors; drug delivery; electrolysis micropump;
Conference_Titel :
Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS), 2015 Transducers - 2015 18th International Conference on
Conference_Location :
Anchorage, AK
DOI :
10.1109/TRANSDUCERS.2015.7181105