• DocumentCode
    731852
  • Title

    A wireless implantable drug infusion system with integrated dosing sensors

  • Author

    Sheybani, R. ; Meng, E.

  • Author_Institution
    Biomed. Microsyst. Lab., Univ. of Southern California, Los Angeles, CA, USA
  • fYear
    2015
  • fDate
    21-25 June 2015
  • Firstpage
    1045
  • Lastpage
    1048
  • Abstract
    A wireless electrolysis-based drug delivery system with integrated electrochemical dosing sensors suitable for management of chronic conditions is presented. Repeatable delivery performance for identical and variable bolus volumes and flow rates of phosphate buffer saline (PBS) was demonstrated. Successful delivery and sensing of 17.75 ± 0.48 μL boluses (1.78% of reservoir fill volume) of lidocaine was achieved.
  • Keywords
    biomedical equipment; drug delivery systems; electrochemical sensors; electrolysis; organic compounds; bolus volumes; chronic condition management; integrated electrochemical dosing sensors; lidocaine; phosphate buffer saline flow rates; wireless electrolysis-based drug delivery system; wireless implantable drug infusion system; Drug delivery; Drugs; Micropumps; Reservoirs; Sensors; Wireless sensor networks; Wires; Electrochemical sensors; drug delivery; electrolysis micropump;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS), 2015 Transducers - 2015 18th International Conference on
  • Conference_Location
    Anchorage, AK
  • Type

    conf

  • DOI
    10.1109/TRANSDUCERS.2015.7181105
  • Filename
    7181105