DocumentCode :
73641
Title :
Moveable Wire Electrode Microchamber for Nanosecond Pulsed Electric-Field Delivery
Author :
Yu-Hsuan Wu ; Arnaud-Cormos, D. ; Casciola, M. ; Sanders, J.M. ; Leveque, P. ; Vernier, P.T.
Author_Institution :
Mork Family Dept. of Chem. Eng. & Mater. Sci., Univ. of Southern California, Los Angeles, CA, USA
Volume :
60
Issue :
2
fYear :
2013
fDate :
Feb. 2013
Firstpage :
489
Lastpage :
496
Abstract :
In this paper, an electromagnetic characterization of a moveable wire electrode microchamber for nanosecond pulse delivery is proposed. The characterization of the exposure system was carried out through experimental measurements and numerical simulations. The frequency and time domain analyses demonstrate the utility of the proposed assembly for delivering pulses as short as 2.5 ns. High-voltage measurements (~1.2 kV) were also performed using pulse generators based on two different technologies with applied pulse durations of 5.0 and 2.5 ns. Validation of the delivery system was accomplished with biological experiments involving cell electroporation with 2.5 and 5.0 ns, 10-MV/m pulsed electric fields. A dose-dependent area increase (osmotic swelling) of the Jurkat cells was observed with pulses as short as 2.5 ns.
Keywords :
bioelectric phenomena; biomedical electrodes; blood; cancer; cellular biophysics; microelectrodes; numerical analysis; osmosis; swelling; time-frequency analysis; voltage measurement; Jurkat cells; applied pulse durations; biological experiments; cell electroporation; dose-dependent area; electromagnetic characterization; experimental measurements; exposure system; frequency-time domain analysis; high-voltage measurements; moveable wire electrode microchamber; nanosecond pulsed electric-field delivery; numerical simulations; osmotic swelling; pulse generators; time 2.5 ns; time 5.0 ns; Electric fields; Electrodes; Generators; Impedance; Tungsten; USA Councils; Wires; Cell electroporation; finite difference time domain (FDTD); high voltage; nanosecond pulsed electric field (nsPEF); wire electrodes; Cell Size; Computer Simulation; Electrodes; Electromagnetic Fields; Electroporation; Humans; Jurkat Cells; Microtechnology; Tungsten;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2012.2228650