• DocumentCode
    739206
  • Title

    The System Design and Evaluation of a 7-DOF Image-Guided Venipuncture Robot

  • Author

    Balter, Max L. ; Chen, Alvin I. ; Maguire, Timothy J. ; Yarmush, Martin L.

  • Author_Institution
    Dept. of Biomed. Eng., Rutgers Univ., Piscataway, NJ, USA
  • Volume
    31
  • Issue
    4
  • fYear
    2015
  • Firstpage
    1044
  • Lastpage
    1053
  • Abstract
    Accessing the venous bloodstream to deliver fluids or obtain a blood sample is the most common clinical routine practiced in the U.S. Practitioners continue to rely on manual venipuncture techniques, but success rates are heavily dependent on clinician skill and patient physiology. In the U.S., failure rates can be as high as 50% in difficult patients, making venipuncture the leading cause of medical injury. To improve the rate of first-stick success, we have developed a portable autonomous venipuncture device that robotically servos a needle into a suitable vein under image guidance. The device operates in real time, combining near-infrared and ultrasound imaging, image analysis, and a 7-degree-of-freedom (DOF) robotic system to perform the venipuncture. The robot consists of a 3-DOF gantry to image the patient´s peripheral forearm veins and a miniaturized 4-DOF serial arm to guide the cannula into the selected vein under closed-loop control. In this paper, we present the system architecture of the robot and evaluate the accuracy and precision through tracking, free-space positioning, and in vitro phantom cannulation experiments. The results demonstrate submillimeter accuracy throughout the operating workspace of the manipulator and a high rate of success when cannulating phantom veins in a skin-mimicking tissue model.
  • Keywords
    blood vessels; closed loop systems; design engineering; infrared imaging; manipulators; medical image processing; medical robotics; robot vision; ultrasonic imaging; 3-DOF gantry; 7-DOF image-guided venipuncture robot evaluation; 7-DOF image-guided venipuncture robot system design; cannulating phantom veins; clinician skill; closed-loop control; degree-of-freedom; free-space positioning; image analysis; in vitro phantom cannulation experiments; manipulator; miniaturized 4-DOF serial arm; near-infrared imaging; patient peripheral forearm veins; patient physiology; portable autonomous venipuncture device; skin-mimicking tissue model; ultrasound imaging; venous bloodstream; Imaging; Kinematics; Manipulators; Needles; Robot kinematics; Veins; Computer vision; kinematics; mechanism design; medical robots and systems; visual servoing;
  • fLanguage
    English
  • Journal_Title
    Robotics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1552-3098
  • Type

    jour

  • DOI
    10.1109/TRO.2015.2452776
  • Filename
    7169572