Study on Further Modification of Non-Contact Capacitance Type-Level Transducer for a Conducting Liquid
Author :
Chakraborty, Sudipta ; Bera, Saikat Kumar ; Mandal, Nirupama ; Bera, Satish Chandra
Author_Institution :
Department of Electronics Engineering, Indian School of Mines, Dhanbad, India
Volume :
15
Issue :
11
fYear :
2015
Firstpage :
6678
Lastpage :
6688
Abstract :
A non-contact capacitance type liquid-level transducer for a conducting liquid consists of a short circuited non-inductive coil wound on a uniform cylinder made of insulating material, such as glass, PVC, nylon, and teflon. The cylinder is connected with a storage tank through a metallic connector and the capacitance between upper short circuited end of the coil and metallic connector varies linearly with liquid level measured from datum level. The capacitance at datum level depends on atmospheric condition and the fringe effect of other conductors. So, the transducers require frequent recalibration depending on atmospheric and environmental conditions. In the present paper, this non-contact capacitive sensor has been further modified in order to minimize this defect of non-contact capacitive sensor. Furthermore, all the conventional capacitance type-level sensors are associated with a perpendicular plate capacitance between liquid and electrode, which varies with the length of the electrode. In this paper, the effect of this perpendicular capacitor on liquid level measured has been studied and its effect is minimized. The proposed modified transducer has been designed and developed and its theory of operation has been derived. The performance of the developed transducer has been experimentally studied and the experimental results are reported in the paper. A good linear performance characteristic of the transducer with very minimum parasitic capacitance effect has been observed.
Keywords :
Capacitance; Capacitive sensors; Electron tubes; Liquids; Storage tanks; Transducers; Conducting liquid; Differential capacitance measurement; Liquid level transducer; Noninductive coil; Op-amp; Perpendicular plate capacitance; conducting liquid; differential capacitance measurement; non-inductive coil; op-amp; perpendicular plate capacitance;
fLanguage :
English
Journal_Title :
Sensors Journal, IEEE
Publisher :
ieee
ISSN :
1530-437X
Type :
jour
DOI :
10.1109/JSEN.2015.2464072
Filename :
7174935
Link To Document :
بازگشت