Characteristics of a PDP-Based Radiation Detector Dependent on Different Electrode Structures
Author :
Eom, Sangheum ; Kim, Jongseok ; Kang, Jungwon
Author_Institution :
Department of Electronics and Electrical Engineering, Dankook University, Yongin, Korea
Volume :
43
Issue :
9
fYear :
2015
Firstpage :
3200
Lastpage :
3204
Abstract :
In this study, a flat-panel detector based on plasma display technology was investigated as a candidate for a flat-panel radiation detector. We studied the dependence of multiplication factors on various electrode structures using the 3-D Garfield code that calculates the passage of particles through the gas-filled gap. Prototype detectors having three different electrode structures were designed and fabricated based on the simulation results. The performances of these detectors were examined by measuring the collected charge density, dark current density, and sensitivity. The collected charge density had the highest value at the condition when the ridged electrode structure was 1.57 \\mu text{C} /cm2 at −1500 V. The dark current of the same detector was varied between 4.8 and 6.17 nA/cm2 at the bias range of −500 to −1500 V. The sensitivity of the ridged electrode detector was 0.363 nC/ {\\rm mR},\\cdot ,{\\rm cm}^{2} at 0.54 V/ \\mu text{m} , and it is approaching to 18% of the sensitivity reported for the commercially available amorphous selenium (a-Se) detector at 10 V/ \\mu text{m} .
Keywords :
Cathodes; Dark current; Detectors; Radiation detectors; Sensitivity; X-ray imaging; Gas-filled detectors; X-ray; X-ray detector; X-ray detector.; radiation detectors; sensors;
fLanguage :
English
Journal_Title :
Plasma Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0093-3813
Type :
jour
DOI :
10.1109/TPS.2015.2459757
Filename :
7177113
Link To Document :
بازگشت