Title :
The Manumeter: A Wearable Device for Monitoring Daily Use of the Wrist and Fingers
Author :
Friedman, Nizan ; Rowe, Justin B. ; Reinkensmeyer, David J. ; Bachman, Mark
Author_Institution :
Dept. of Biomed. Eng., Univ. of California, Irvine, Irvine, CA, USA
Abstract :
Nonobtrusive options for monitoring the wrist and hand movement are needed for stroke rehabilitation and other applications. This paper describes the “manumeter,” a device that logs total angular distance travelled by wrist and finger joints using a magnetic ring worn on the index finger and two triaxial magnetometers mounted in a watch-like unit. We describe an approach to estimate the wrist and finger joint angles using a radial basis function network that maps differential magnetometer readings to joint angles. We tested this approach by comparing manumeter estimates of total angular excursion with those from a passive goniometric exoskeleton worn simultaneously as seven participants completed a set of 12 manual tasks at low-, medium-, and high-intensity conditions on a first testing day, 1-2 days later, and 6-8 days later, using only the original calibration from the first testing day. Manumeter estimates scaled proportionally to the intensity of hand activity. Estimates of angular excursion made with the manumeter were 92.5% ± 28.4 (SD), 98.3% ± 23.3, and 94.7% ± 19.3 of the goniometric exoskeleton across the three testing days, respectively. Magnetic sensing of wrist and finger movement is nonobtrusive and can quantify the amount of use of the hand across days.
Keywords :
biomechanics; biomedical measurement; body sensor networks; calibration; goniometers; magnetometers; patient rehabilitation; radial basis function networks; calibration; differential magnetometer; finger joint angle estimation; finger joint movement monitoring; hand movement monitoring; manumeter device; passive goniometric exoskeleton; radial basis function network; stroke rehabilitation; time 1 day to 2 day; time 6 day to 8 day; triaxial magnetometers; wearable device; wrist joint angle estimation; wrist joint movement monitoring; Accelerometers; Biomedical monitoring; Fingers; Magnetic field measurement; Magnetometers; Patient monitoring; Patient rehabilitation; Sensors; Wearable computers; Wrist; Accelerometers; arm monitoring; biomedical monitoring; hand monitoring; magnetic field measurement; magnetometers; patient monitoring; patient rehabilitation; sensors; wearable;
Journal_Title :
Biomedical and Health Informatics, IEEE Journal of
DOI :
10.1109/JBHI.2014.2329841