Title : 
High speed pulse digitization in driftless gas scintillation proportional counters
         
        
            Author : 
Valero, J. ; Smith, A. ; Peacock, A.
         
        
            Author_Institution : 
ESA, Noordwijk, Netherlands
         
        
        
        
        
            fDate : 
6/1/1988 12:00:00 AM
         
        
        
        
            Abstract : 
The experimental arrangement used to examine light pulses of the driftless-gas-scintillation proportional counter are described and two applications using pulse-shape parameters are discussed. Since true X-rays should have a profile at least similar to a particular shape, events with a widely different profile can be rejected immediately. Events that have no software-discernable plateau are labeled spurious. This procedure typically rejects 60% of all laboratory background events. The problem of the dependency of the integrated light pulse (pulse height) on burst length is the second application considered for use with higher-energy X-rays
         
        
            Keywords : 
X-ray detection and measurement; proportional counters; scintillation counters; X-rays; burst length; driftless gas scintillation proportional counters; integrated light pulse; light pulses; pulse height; pulse-shape parameters; Anodes; Clouds; Electrons; Energy resolution; Instruments; Photonics; Pulse amplifiers; Radiation detectors; Shape; Solid scintillation detectors;
         
        
        
            Journal_Title : 
Nuclear Science, IEEE Transactions on