• DocumentCode
    747660
  • Title

    Determination of DC voltage ratio of a self-calibrating DC voltage divider

  • Author

    Liu, Ling Xiang ; Chua, Sze Wey ; Ang, Chee Kiang

  • Author_Institution
    Electr. Metrol. Dept., Nat. Metrol. Center, Singapore, Singapore
  • Volume
    54
  • Issue
    2
  • fYear
    2005
  • fDate
    4/1/2005 12:00:00 AM
  • Firstpage
    571
  • Lastpage
    575
  • Abstract
    This paper presents two techniques for determining the ratios of resistive dc voltage dividers up to 1 kV. The techniques are based on absolute calibration methods. The setup requires equipment commonly available in most standard laboratories and is capable of providing high-accuracy calibration under practical laboratory working conditions.
  • Keywords
    calibration; constants; laboratory techniques; voltage dividers; absolute calibration; dc voltage ratio; high-accuracy calibration; laboratory working conditions; resistive dc voltage dividers; self-calibrating dc voltage divider; Calibration; Circuit testing; Detectors; Employee welfare; Laboratories; Measurement standards; Metrology; Productivity; Switches; Voltage measurement; Calibration; dc voltage; dc voltage ratio; self-calibrating divider;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2004.843089
  • Filename
    1408236