• DocumentCode
    751867
  • Title

    A high-speed data acquisition system for ECT based on the differential sampling method

  • Author

    Wang, Baoliang ; Ji, Haifeng ; Huang, Zhiyao ; Li, Haiqing

  • Author_Institution
    Dept. of Control Sci. & Eng., Zhejiang Univ., Hangzhou, China
  • Volume
    5
  • Issue
    2
  • fYear
    2005
  • fDate
    4/1/2005 12:00:00 AM
  • Firstpage
    308
  • Lastpage
    312
  • Abstract
    A new capacitance measuring circuit based on differential sampling methods for electrical capacitance tomography has been developed. The measured capacitance is only charged and discharged once and no noise is created by the measurement principle. A capacitance measurement system for 12 electrodes is constructed based on the new measurement circuit. Test results show the data acquisition time for 66 capacitances (for reconstructing one image) is less than 1.25 ms.
  • Keywords
    capacitance measurement; data acquisition; measurement systems; tomography; capacitance measurement system; differential sampling methods; electrical capacitance tomography; high-speed data acquisition system; measurement circuit; sampling method; Capacitance measurement; Circuit noise; Circuit testing; Current measurement; Data acquisition; Electric variables measurement; Electrical capacitance tomography; Electrodes; Noise measurement; Sampling methods; Capacitance measurement; tomography;
  • fLanguage
    English
  • Journal_Title
    Sensors Journal, IEEE
  • Publisher
    ieee
  • ISSN
    1530-437X
  • Type

    jour

  • DOI
    10.1109/JSEN.2004.842627
  • Filename
    1411811