• DocumentCode
    752533
  • Title

    An automated system to control the polarization voltage of silicon detectors

  • Author

    Altieri, S. ; Fossati, F. ; Lanza, A. ; Pinelli, T.

  • Author_Institution
    Dipartimento di Fisica Nucl. e Teorica, Istituto Nazionale di Fisica Nucl., Pavia, Italy
  • Volume
    42
  • Issue
    2
  • fYear
    1995
  • fDate
    4/1/1995 12:00:00 AM
  • Firstpage
    57
  • Lastpage
    60
  • Abstract
    A remote controlled system to maintain polarization voltage across a silicon detector within a user defined window is described. The system regulates silicon bias voltage, caused by increased leakage current due to radiation damage, by switching external resistors to maintain a constant voltage drop across the bias load resistor, independent of leakage current. This Z80 microprocessor based system links with the host computer by a RS-232 serial port. The voltage resolution of the system is less than 0.2 V in the worst case
  • Keywords
    counter accessories; high energy physics instrumentation computing; radiation effects; silicon radiation detectors; 0.2 V; RS-232 serial port; Si bias voltage; Si detectors; Z80 microprocessor based system; automated system; bias load resistor; constant voltage drop; external resistors; host computer; leakage current; polarization voltage; radiation damage; remote controlled system; user defined window; voltage resolution; Automatic voltage control; Circuits; Control systems; Delay; Detectors; Leakage current; Polarization; Resistors; Silicon; Voltage control;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.372132
  • Filename