• DocumentCode
    754903
  • Title

    Application of adaptive filters to noninvasive acoustical detection of coronary occlusions before and after angioplasty

  • Author

    Akay, Metin ; Akay, Yasemin M. ; Welkowitz, Walter ; Semmlow, John L. ; Kostis, John B.

  • Author_Institution
    Dept. of Biomed. Eng., Rutgers Univ., Piscataway, NJ, USA
  • Volume
    39
  • Issue
    2
  • fYear
    1992
  • Firstpage
    176
  • Lastpage
    184
  • Abstract
    Isolated diastolic heart sounds taken from recordings made at the patient´s bedside were modeled using the autoregressive (AR) and autoregressive moving average (ARMA) methods after adaptive line enhancement (ALE). Decisions were made in a blind fashion without prior knowledge of whether a given recording was made before or after angioplasty. Resulting model frequency spectra showed greater high frequency components (between 400 and 800 Hz) in preangioplasty patients, and a consistent shift in amplitude of the second pole pairs of the AR and ARMA methods with surgery. Blind assessment based on frequency spectra and poles, correctly classified the diastolic recordings in 18 of 20 cases. These results provide evidence supporting the hypothesis that coronary stenoses produce detectable sounds during diastole.
  • Keywords
    adaptive filters; bioacoustics; cardiology; patient diagnosis; 400 to 800 Hz; adaptive filters; adaptive line enhancement; angioplasty; autoregressive moving average; coronary occlusions; coronary stenoses; high frequency components; isolated diastolic heart sounds; medical recordings; model frequency spectra; noninvasive acoustical detection; pole pairs; preangioplasty patients; surgery; Acoustic signal detection; Adaptive filters; Angioplasty; Arteries; Autoregressive processes; Blood flow; Coronary arteriosclerosis; Heart; Myocardium; Testing; Angioplasty, Transluminal, Percutaneous Coronary; Coronary Disease; Filtration; Heart Auscultation; Heart Sounds; Humans; Image Enhancement; Likelihood Functions; Myocardial Contraction; Rheology; Sensitivity and Specificity; Signal Processing, Computer-Assisted; Single-Blind Method;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.121649
  • Filename
    121649