  • Abstract
    Can cellular phones and personal communication systems base station antennas affect the active or passive implantable medical devices adversely? Concerns over the possible harmful effects of nonionizing irradiation upon implanted medical devices have been present for many years. Key issues to address are the questions of whether mobile phones have a detrimental effect on implants, and how the interaction of the handset with the body can be minimized in order to both alleviate public fears and improve handset antenna performance and new implant designs. This paper presents a thorough investigation of the scattering of an electromagnetic (EM) wave from a perfectly conducting implant (a cylindrical wire and a very thin cylindrical disk) of electrically small radius (of resonant length), embedded eccentrically into a dielectric spherical head model by a dipole antenna (0.4 wavelength) at 900 MHz. The dyadic Green´s function (DGF) for spherical vector wave functions is employed. Analytical expressions for the scattered fields of an implant embedded head model is obtained. Numerical results from analytical expressions are computed for this problem and then compared with the results from the same model using the finite-difference time-domain, EMU-FDTD electromagnetic simulator. Good agreement is observed between the analytical results on the proposed method in comparison with the FDTD method
  • Keywords
    Green´s function methods; biological effects of fields; cellular radio; finite difference time-domain analysis; prosthetics; 900 MHz; EMU-FDTD electromagnetic simulator; cellular phones; dipole antenna; dyadic Green function; electromagnetic field effects; electromagnetic wave scattering; finite-difference time-domain analysis; implantable medical devices; implanted dielectric spherical head model; mobile communication frequency; personal communication systems base station antennas; spherical vector wave functions; Cellular phones; Electromagnetic fields; Electromagnetic modeling; Electromagnetic scattering; Finite difference methods; Frequency; Head; Implants; Mobile communication; Telephone sets; Cylinder; dipole; dyadic Green´s function (DGF); electromagnetic (EM); head; implant; mobile phone; spheres; Cellular Phone; Computer Simulation; Electromagnetic Fields; Environmental Exposure; Head; Humans; Microwaves; Models, Biological; Prostheses and Implants; Radiation Dosage; Radiometry; Relative Biological Effectiveness;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2006.881770
  • Filename
    1703761