Title :
Data transmission from an implantable biotelemeter by load-shift keying using circuit configuration modulator
Author :
Tang, Zhengnian ; Smith, Brian ; Schild, John H. ; Peckham, P. Hunter
Author_Institution :
Dept. of Biomed. Eng., Case Western Reserve Univ., Cleveland, OH, USA
fDate :
5/1/1995 12:00:00 AM
Abstract :
Using the reflected impedance property of an inductive couple (transformer), a modulation method, load-shift keying using circuit configuration modulator (LSK-CCM), was developed to perform data transmission from an implantable telemeter. With a very simple circuit, this method utilizes a radio-frequency electromagnetic field induced with a single pair of coils to transmit power into the implant and data out of it
Keywords :
biological techniques; biomedical electronics; biomedical telemetry; modulation; circuit configuration modulator; coil pair; data transmission; implantable biotelemeter; inductive couple; load-shift keying; power transmission; radio-frequency electromagnetic field induction; reflected impedance property; simple circuit; transformer; Amplitude shift keying; Coils; Coupling circuits; Data communication; Digital modulation; Frequency modulation; Impedance; Implants; Radio frequency; Telemetry;
Journal_Title :
Biomedical Engineering, IEEE Transactions on