DocumentCode
761164
Title
Vector characterization of photodetectors, photoreceivers, and optical pulse sources by time-domain pulse response measurements
Author
Hawkins, R. Thomas, II ; Jones, Michael D. ; Pepper, Steven H. ; Goll, Jeffrey H. ; Ravel, Mihir K.
Author_Institution
Tektronix Inc., Beaverton, OR, USA
Volume
41
Issue
4
fYear
1992
fDate
8/1/1992 12:00:00 AM
Firstpage
467
Lastpage
475
Abstract
Deconvolution of measurement system effects from pulse response measurements is demonstrated to yield reproducible, accurate characterization of the impulse response (and vector frequency response) of a photodetector or photoreceiver, as well as the intensity waveform of an optical pulse. Calibration is based on a <3-ps FWHM (full width at half maximum) optical pulse and a 50-GHz 3-dB bandwidth electrical sampling system. Vector characterization of a photodetector/photoreceiver to >40 GHz and an optical pulse source to >30 GHz are demonstrated. Calibration and effects of noise are discussed
Keywords
calibration; optical variables measurement; photodetectors; waveform analysis; 30 GHz; 40 GHz; 50 GHz; FFT; impulse response; intensity waveform; noise; optical pulse sources; oscilloscope; photodetectors; photoreceivers; pulse response measurements; time-domain pulse response measurements; vector frequency response; Bandwidth; Calibration; Deconvolution; Frequency measurement; Frequency response; Optical noise; Optical pulses; Photodetectors; Pulse measurements; Sampling methods;
fLanguage
English