DocumentCode :
761316
Title :
Upscaling DC resistance using a Josephson junction array
Author :
Kohlmann, Johannes ; Gutmann, Peter ; Niemeyer, Jurgen
Author_Institution :
Phys.-Tech. Bundesanstalt, Braunschweig, Germany
Volume :
41
Issue :
4
fYear :
1992
fDate :
8/1/1992 12:00:00 AM
Firstpage :
548
Lastpage :
551
Abstract :
A ratio standard for DC resistance measurements in the range above about 10 kΩ using a three-terminal Josephson junction series array is described. Resistance ratios of two-terminal resistors from 1:1 to at least 1:10 are examined with a single device utilizing the voltage steps of the Josephson array as voltage sources of a potentiometer bridge configuration. Some features of the circuit are discussed. The system promises systematic ratio errors below 10-8 and uncertainties which depend only on the quality of the resistors. Preliminary measurements with a prototype instrument show a relative reproducibility of 1×10-7, limited by the noise of a commercial digital multimeter used as a detector
Keywords :
Josephson effect; bridge instruments; electric resistance measurement; electric sensing devices; superconducting junction devices; DC resistance measurements; digital multimeter; measurement errors; noise; potentiometer bridge; prototype instrument; ratio standard; reproducibility; three-terminal Josephson junction series array; two-terminal resistors; voltage sources; voltage steps; Bridge circuits; Electrical resistance measurement; Instruments; Josephson junctions; Measurement standards; Noise measurement; Potentiometers; Prototypes; Resistors; Voltage;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/19.155923
Filename :
155923
Link To Document :
بازگشت