DocumentCode :
76405
Title :
Characterization of a Wireless Potentiostat for Integration With a Novel Implantable Biotransducer
Author :
Kotanen, Christian N. ; Guiseppi-Elie, A.
Author_Institution :
Dept. of Bioeng., Clemson Univ., Anderson, SC, USA
Volume :
14
Issue :
3
fYear :
2014
fDate :
Mar-14
Firstpage :
768
Lastpage :
776
Abstract :
Wireless potentiostats are being developed and commercialized for use in the development of implantable electrochemical biosensors for the monitoring of physiological markers in a wide range of pathologies. The merits and drawbacks of the Pinnacle 8151 dual potentiostat were investigated for use with a novel implantable biotransducer, the dual responsive microdisc electrode array (MDEA 5037) of ABTECH Scientific. The potentiostat was evaluated with R and RC dummy cells and with MDEA transducers modified with polypyrrole-poly(2-hydroxyethyl methacrylate) electroconductive hydrogels that were all qualified by calibrated ac impedance spectroscopy. In a laboratory setting, the 8151 potentiostat maintained a steady and unbroken signal as far away as 75 ft (23 m), and when unobstructed, up to 100 ft (30 m). The error between the two channels was determined to be 6.4 (±0.3)% and 0.7 (±0.1)% for resistor-capacitor dummy cells ( RC=nominally 10 MΩ and 1 μF) and resistor dummy cells (R=nominally 10 MΩ), respectively. The intra-channel variability was judged to be too large for exacting analyte determinations.
Keywords :
biomedical electrodes; biomedical transducers; biosensors; calibration; capacitors; diseases; electrochemical electrodes; electrochemical sensors; hydrogels; prosthetics; resistors; ABTECH scientific; MDEA transducers; R cells; RC dummy cells; calibrated ac impedance spectroscopy; dual responsive microdisc electrode array; implantable biotransducer; implantable electrochemical biosensors; intrachannel variability; laboratory setting; physiological marker monitoring; polypyrrole-poly(2-hydroxyethyl methacrylate) electroconductive hydrogels; resistor-capacitor dummy cells; wireless potentiostat characterization; Biosensors; Electric potential; Electrodes; Radiation detectors; Resistors; Transducers; Wireless communication; Biomedical telemetry; biomedical transducers; biosensors; polypyrrole; telemedicine;
fLanguage :
English
Journal_Title :
Sensors Journal, IEEE
Publisher :
ieee
ISSN :
1530-437X
Type :
jour
DOI :
10.1109/JSEN.2013.2288059
Filename :
6651733
Link To Document :
بازگشت