DocumentCode :
764096
Title :
Differential method against motion artefacts in skin conductance response measurement
Author :
Lee, G. ; Choi, J.
Author_Institution :
Sch. of Eng., Inf. & Commun. Univ., Daejeon
Volume :
43
Issue :
7
fYear :
2007
Firstpage :
375
Lastpage :
377
Abstract :
A differential method is proposed that can reduce motion artefacts in the bipolar measurement of skin conductance response. It is based on the fact that both the conductive and the capacitive currents across the skin are proportional to the contact area while only the conductive current reflects the sweat gland activity. Experimental results with a prototype system are presented
Keywords :
bioelectric potentials; biomedical measurement; electric admittance measurement; skin; bipolar measurement; capacitive currents; conductive currents; contact area; motion artefacts; skin conductance response measurement; sweat gland activity;
fLanguage :
English
Journal_Title :
Electronics Letters
Publisher :
iet
ISSN :
0013-5194
Type :
jour
DOI :
10.1049/el:20073717
Filename :
4145283