Title :
Mechanocardiograph Due to Amorphous Wire-Core Multivibrator Magnetometer
Author :
Mohri, K. ; Kondo, T.
Author_Institution :
Kyushu Institute of Technology, Kitakyushu.
Abstract :
Application of a noncontact displacement sensor, consisting of a small magnet and amorphous dual-core magnetometer and featuring excellent linearity, to a mechanocardiograph (heart displacement sensor) is described. Detection signals showed linear dependence on magnet displacement (nonlinearity ≪ 1%) for displacements up to 6 mm. The minimum detection resolution was 1 ¿m, enabling detection of human heart displacement, which ranges from a few to several tens of ¿m. Cardiac displacement distributions can be obtained, for more complete diagnostics than are possible with electroor magnetocardiographs.
Keywords :
Amorphous magnetic materials; Amorphous materials; Heart; Humans; Linearity; Magnetic sensors; Magnetometers; Sensor phenomena and characterization; Signal detection; Signal resolution;
Journal_Title :
Magnetics in Japan, IEEE Translation Journal on
DOI :