Title : 
Cross-conductivity standard for nonferrous metals
         
        
            Author : 
He, Chaolai ; Lu, Zuliang ; Liu, Songyin ; Liu, Ruimim
         
        
            Author_Institution : 
Nat. Inst. of Metrol., Beijing, China
         
        
        
        
        
            fDate : 
4/1/1995 12:00:00 AM
         
        
        
        
            Abstract : 
A novel cross-conductivity standard based on a new principle developed from the Thompson-Lampard Theorem has been established at National Institute of Metrology. The conductivity range of the standard is approximately from 3%IACS to 100% IACS with an uncertainty of about 0.07%. The unknowns can be calibrated with an uncertainty of (0.1%rdg+0.12% IACS) by using an ac measuring system and this new method
         
        
            Keywords : 
Thomson effect; electrical conductivity measurement; measurement standards; National Institute of Metrology; Thompson-Lampard theorem; absolute measurement; ac measuring system; cross-conductivity standard; dc measurement; nonferrous metals; uncertainty; Area measurement; Conductivity measurement; Conductors; Current measurement; Electrostatic measurements; Helium; Measurement standards; Metrology; Thickness measurement; Voltage;
         
        
        
            Journal_Title : 
Instrumentation and Measurement, IEEE Transactions on