DocumentCode
766882
Title
Electromagnetic Determination of Carotid Blood Flow in the Anesthetized Rat
Author
Cox, P. ; Arora, H. ; Kolin, A.
Author_Institution
Department of Biophysics and Nuclear Medicine, University of California, Los Angeles, Calif.
Volume
10
Issue
4
fYear
1963
Firstpage
171
Lastpage
173
Abstract
The paper describes the design and performance of the smallest iron core electromagnetic flow transducer built so far. The design is miniaturized as to the total volume of the device as well as to the size of the artery it accommodates. The use of this transducer for recording of blood flow in the smallest species of animal employed until now in blood flow research is described. Illustrations of pharmacological observations in anesthetized rats are presented.
Keywords
Animals; Arteries; Blood flow; Electrodes; Magnetic shielding; Perpendicular magnetic recording; Rats; Toroidal magnetic fields; Transducers; Wiring;
fLanguage
English
Journal_Title
Bio-medical Electronics, IEEE Transactions on
Publisher
ieee
ISSN
0096-0616
Type
jour
DOI
10.1109/TBMEL.1963.4322827
Filename
4322827
Link To Document