Title : 
Calibration of RF field strength meters
         
        
            Author : 
de Vreede, J.P.M. ; Teunisse, G.M. ; Munter, K. ; Hoitink, H. ; Holland, K. ; O´Dwyer, M. ; Bergsten, L. ; Puranen, L.
         
        
            Author_Institution : 
NMi Van Swinden Lab., Delft, Netherlands
         
        
        
        
        
            fDate : 
4/1/1995 12:00:00 AM
         
        
        
        
            Abstract : 
An intercomparison of field strength meters was performed during 1991-1993 over the frequency range 100 kHz-1000 MHz and field strength range 1-10 V/m using different calibration techniques. Four field sensors, electric as well as magnetic, have been used. Seven institutes participated in this intercomparison using at least three different calibration methods. A calibration uncertainty of less than 1 dB can be obtained if certain procedures are followed. The results are applicable in the fields of EMC testing and health care (nonionizing radiation)
         
        
            Keywords : 
calibration; electromagnetic compatibility; field strength measurement; health hazards; radiation protection; radiation therapy; 100 kHz to 1000 MHz; EMC testing; RF field strength meters; calibration; health care; intercomparison; nonionizing radiation; uncertainty; Calibration; Electromagnetic compatibility; Immunity testing; Laboratories; Magnetic sensors; Measurement standards; Metrology; Probes; Radiation safety; Radio frequency;
         
        
        
            Journal_Title : 
Instrumentation and Measurement, IEEE Transactions on