• DocumentCode
    767026
  • Title

    Very low frequency ac-dc transfer standard

  • Author

    Budovsky, Ilya

  • Author_Institution
    Div. of Appl. Phys., CSIRO, Lindfield, NSW, Australia
  • Volume
    44
  • Issue
    2
  • fYear
    1995
  • fDate
    4/1/1995 12:00:00 AM
  • Firstpage
    367
  • Lastpage
    369
  • Abstract
    A low frequency ac-dc transfer standard has been developed to satisfy the NML demand for very low frequency ac-dc transfer calibrations. Tests performed using digitally synthesized signals and synchronised dc measurements showed less than 13 parts per million variation in the ac-dc difference of the standard within the frequency range 0.01 to 100 Hz
  • Keywords
    calibration; electric current measurement; measurement standards; voltage measurement; 0.01 to 100 Hz; NML; ac-dc difference; calibration; digitally synthesized signals; low frequency ac-dc transfer standard; synchronised dc measurements; Analog-digital conversion; Calibration; Frequency conversion; Frequency synchronization; Measurement standards; Performance evaluation; Standards development; Temperature; Testing; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.377854
  • Filename
    377854