Abstract :
In order to improve amenity for patients in hospitals, we have designed an open type of magnetically shielded room (MSR) for magnetic resonance imaging combined with cancelling coils, which enable transparent opening. First, an optimal arrangement of cancelling coils is investigated using three-dimensional magnetic field analysis. Secondly, the validity of the analysis method is verified and the realization of an open type of MSR designed both for keeping low leakage flux density and for maximizing the opening is confirmed by experiments using a small MSR model
Keywords :
biomedical MRI; magnetic fields; magnetic shielding; patient diagnosis; 3D magnetic field analysis; cancelling coil; flux density; magnetic resonance imaging; magnetically shielded room; patient diagnosis; transparent opening; Bars; Coils; Computed tomography; Hospitals; Magnetic analysis; Magnetic field measurement; Magnetic flux; Magnetic flux leakage; Magnetic resonance imaging; Magnetic shielding; Amenity of patients; cancelling coil; magnetic field analysis; magnetic resonance imaging; magnetically shielded room;