Title :
Open Type of Magnetically Shielded Room Combined With Cancelling Coil for Magnetic Resonance Imaging
Author :
Hirosato, Shigetaka ; Yamazaki, Keita ; Muramatsu, Kazuhiro ; Hirayama, Masayuki ; Haga, Akira ; Katada, Kazuhiro
Author_Institution :
R & D Inst., Takenaka Corp
Volume :
42
Issue :
10
fYear :
2006
Firstpage :
3542
Lastpage :
3544
Abstract :
In order to improve amenity for patients in hospitals, we have designed an open type of magnetically shielded room (MSR) for magnetic resonance imaging combined with cancelling coils, which enable transparent opening. First, an optimal arrangement of cancelling coils is investigated using three-dimensional magnetic field analysis. Secondly, the validity of the analysis method is verified and the realization of an open type of MSR designed both for keeping low leakage flux density and for maximizing the opening is confirmed by experiments using a small MSR model
Keywords :
biomedical MRI; magnetic fields; magnetic shielding; patient diagnosis; 3D magnetic field analysis; cancelling coil; flux density; magnetic resonance imaging; magnetically shielded room; patient diagnosis; transparent opening; Bars; Coils; Computed tomography; Hospitals; Magnetic analysis; Magnetic field measurement; Magnetic flux; Magnetic flux leakage; Magnetic resonance imaging; Magnetic shielding; Amenity of patients; cancelling coil; magnetic field analysis; magnetic resonance imaging; magnetically shielded room;
fLanguage :
English
Journal_Title :
Magnetics, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9464
Type :
jour
DOI :
10.1109/TMAG.2006.883835
Filename :
1704687
Link To Document :
بازگشت