DocumentCode :
768203
Title :
Improvement of the Transcutaneous Energy Transmission System Utilizing Ferrite Cored Coils for Artificial Hearts
Author :
Miura, Hidekazu ; Arai, Shinsuke ; Kakubari, Yasuyuki ; Sato, Fumihiro ; Matsuki, Hidetoshi ; Sato, Tadakuni
Author_Institution :
Graduate Sch. of Eng., Tohoku Univ.
Volume :
42
Issue :
10
fYear :
2006
Firstpage :
3578
Lastpage :
3580
Abstract :
Newly designed transcutaneous energy transmission system for a ventricular assist device was developed. We obtained higher dc-dc efficiency of 93.4% and smaller size of devices. Also, the devices were designed in view of the biomedical compatibility, gradually sloped coils and the flexibility of the primary coil prevent pressure necroses of the skin. The rectifier circuit board was placed in the internal space of the ferrite core to reduce the number of the implanted devices. However, radiated outward ground patterns were adopted to achieve a balance between thermal radiation and reduction of eddy-current loss
Keywords :
artificial organs; coils; ferrites; heat radiation; prosthetic power supplies; artificial hearts; contactless power supply; ferrite cored coils; pressure necroses; rectifier circuit board; synchronous rectification; thermal radiation; transcutaneous energy transmission; Artificial heart; Coils; Ferrites; Flexible printed circuits; Magnetic cores; Magnetic resonance; Phase locked loops; Printed circuits; Rectifiers; Voltage; Artificial heart; contactless power supply; digital phase-locked loop (PLL); synchronous rectification;
fLanguage :
English
Journal_Title :
Magnetics, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9464
Type :
jour
DOI :
10.1109/TMAG.2006.879630
Filename :
1704698
Link To Document :
بازگشت