    Compact readout electronics for position sensitive photomultiplier tubes

  • Author

    Olcott, Peter Demetri ; Talcott, Jonathon A. ; Levin, Craig Steven ; Habte, Frezghi ; Foudray, Angela M K

  • Author_Institution
    Molecular Imaging Program, Stanford Univ. Sch. of Med., CA, USA
  • Volume
    52
  • Issue
    1
  • fYear
    2005
  • Firstpage
    21
  • Lastpage
    27
  • Abstract
    Two charge multiplexed readout schemes for position-sensitive photomultiplier tubes (PSPMTs) have been developed and evaluated. A two-stage split-charge method is compared to standard positional charge division. The resulting electronic circuits are highly compact and will be initially incorporated into a hand-held gamma ray imager for cancer staging. Electrical tests on the readout electronics are performed and intrinsic scintillation camera performance is evaluated using a 23 × 23 array of pixilated 2 × 2 × 3 mm3 LSO crystals coupled to a H8500 flat-panel PSPMT.
  • Keywords
    biomedical imaging; cancer; gamma-ray detection; nuclear electronics; photomultipliers; position sensitive particle detectors; readout electronics; solid scintillation detectors; 2 mm; 3 mm; H8500 flat-panel PSPMT; cancer staging; charge multiplexed readout schemes; compact readout electronics; electrical tests; hand-held gamma ray imager; intrinsic scintillation camera performance; pixilated LSO crystals; position sensitive photomultiplier tubes; standard positional charge division; two-stage split-charge method; Cameras; Cancer; Circuit testing; Crystals; Electron tubes; Electronic circuits; Electronic equipment testing; Performance evaluation; Photomultipliers; Readout electronics;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2004.843134
  • Filename
    1417103