DocumentCode :
773416
Title :
Gain vs Temperature Effects in Nal(Tl) Photomultiplier Scintillation Detectors Using 10 and 14 Stage Tubes
Author :
Rohde, R.Erline
Author_Institution :
Electronics Division Argonne National Laboratory Argonne, Illinois
Volume :
12
Issue :
1
fYear :
1965
Firstpage :
16
Lastpage :
23
Abstract :
The manner in which the gain of a radiation transducer comprised of a 2" × 2" NaI(Tl) crystal, and either a 5-inch 10-stage or 2-inch 14-stage photomultiplier tube viewing a Cs137 source, varies with ambient temperature changes over the range of 0° C to 50° C has been investigated. Gain vs temperature response curves are presented and operating temperatures which minimize gain variations are delineated. A method is given for evaluating the magnitude of gain variations encountered from the curves.
Keywords :
Electron tubes; Gravity; Laboratories; Photomultipliers; Pulse amplifiers; Scintillation counters; Temperature distribution; Temperature sensors; Transducers; Voltage;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1965.4323491