• DocumentCode
    773614
  • Title

    Mechatronic Design of a New Robot for Force Control in Minimally Invasive Surgery

  • Author

    Zemiti, Nabil ; Morel, Guillaume ; Ortmaier, Tobias ; Bonnet, Nicolas

  • Author_Institution
    Lab. de Robotique de Paris, Univ. Pierre et Marie Curie, Paris
  • Volume
    12
  • Issue
    2
  • fYear
    2007
  • fDate
    4/1/2007 12:00:00 AM
  • Firstpage
    143
  • Lastpage
    153
  • Abstract
    Minimally invasive surgery (MIS) challenges the surgeon´s skills due to his/her separation from the operation area, which can be reached with long instruments only. Therefore, the surgeon looses access to the manipulation forces inside the patient. This reduces his/her dexterity when performing the operation. A new compact and lightweight robot for MIS is presented, which allows for the measurement of manipulation forces. The main advantage of this concept is that no miniaturized force sensor has to be integrated into surgical instruments and inserted into the patient. Rather, outside the patient, a standard sensor is attached to a modified trocar, which allows for the undisturbed measurement of manipulation forces. This approach reduces costs and sterilizability demands. Results of in vitro and in vivo force control experiments are presented to validate the concepts
  • Keywords
    force control; manipulators; mechatronics; medical robotics; surgery; force control; mechatronic design; minimally invasive surgery; robot manipulators; surgical instruments; Costs; Force control; Force measurement; Force sensors; Measurement standards; Mechatronics; Minimally invasive surgery; Robot sensing systems; Surges; Surgical instruments; Force control; force measurement; minimally invasive surgery (MIS);
  • fLanguage
    English
  • Journal_Title
    Mechatronics, IEEE/ASME Transactions on
  • Publisher
    ieee
  • ISSN
    1083-4435
  • Type

    jour

  • DOI
    10.1109/TMECH.2007.892831
  • Filename
    4154715