DocumentCode :
776190
Title :
Velocity Spectrometers Used in Bevatron Deflected-Beam Research
Author :
Edwards, George W. ; Watt, Robert D.
Author_Institution :
Lawrence Radiation Laboratory, University of California Berkeley, California
Volume :
12
Issue :
3
fYear :
1965
fDate :
6/1/1965 12:00:00 AM
Firstpage :
922
Lastpage :
924
Abstract :
Two completely redesigned 10-ft, glasscathode, parallel-plate velocity spectrometers (MK V) have been used in particle research at 500 kV for about 3000 hr each without maintenance. Current consumption, including power supply, is approximately 100 ¿A on each plate (< 200 ¿A total). The units do not appear to be gap-sensitive, as one was operated with a 2-in. gap, and the other with a 4-in. gap. No significant operating differences are apparent.
Keywords :
Anodes; Cathodes; Electrodes; Glass; Laboratories; Magnetic fields; Spectroscopy; Steel; Testing; Voltage;