DocumentCode
779738
Title
Encapsulated germanium γ-ray spectrometers performance characteristics and operational experience
Author
Webb, P.P. ; Green, R.M. ; Fowler, I.L. ; Malm, H.L.
Author_Institution
RCA Victor Co. Ltd., Research Laboratories, Montreal, Quebec
Volume
13
Issue
3
fYear
1966
fDate
6/1/1966 12:00:00 AM
Firstpage
351
Lastpage
358
Abstract
Hermetically sealed Ge(Li) diodes for γ-ray spectroscopy have now been manufactured and under study for approximately two years. During this time, ´planar´ drifted encapsulated diodes have been increased in size from 2 cm2 Ã 2 mm deep (0.4 cm3) to ˜ 8 cm2 Ã 10 mm deep (8 cm3). Coaxially drifted diodes of 25 to 30 cm3 volume have also been encapsulated. Encapsulation details of both types of detector are given, together with performance characteristics. Measured intrinsic full-energy peak efficiency curves are given for four detector sizes. For an 8 cm3 planar diode this efficiency is 1.1% at 7 cm for 1 MeV γ-rays. The corresponding efficiency for a 27 cm3 coaxial diode is 3.1%. Voltage ratings of 1000 V/cm with leakage currents ⩽ 1 nA at operating voltage are achieved. Resolutions of 3.4 and 5.5 keV (fwhm) at 1173 keV γ-ray energy have been obtained with a 2 cm3 (5 mm deep) planar diode and a 27 cm3 coaxial detector respectively. Mounting and general application details are discussed.
Keywords
Coaxial components; Detectors; Diodes; Encapsulation; Germanium; Hermetic seals; Manufacturing; Size measurement; Spectroscopy; Voltage;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1966.4324117
Filename
4324117
Link To Document