• DocumentCode
    779911
  • Title

    A Parametric Radiation Detector Preamplifier

  • Author

    Radeka, V. ; Chase, R.L.

  • Author_Institution
    Brookhaven National Laboratory Upton, N. Y.
  • Volume
    13
  • Issue
    3
  • fYear
    1966
  • fDate
    6/1/1966 12:00:00 AM
  • Firstpage
    477
  • Lastpage
    494
  • Abstract
    A charge-sensitive preamplifier has been built employing a pair of varactor diodes in a novel parametric amplifier configuration as the first amplifying element. The circuit consists of a parametric subharmonic oscillator, pumped at about 130 Mc included in a negative feedback loop. Input charge signals perturb the oscillator tuning. The feedback circuit prevents significant change in the oscillator amplitude by supplying an approximately equal, opposite charge to the input circuit through a small feedback capacitor. The following noise performance, referred to germanium detectors, has been obtained. Temperature FWHM Slope 300°K 1.40 keV 20 ev/pF 77°K 0.64 keV 13 ev/pF
  • Keywords
    Capacitors; Circuit optimization; Diodes; Feedback circuits; Negative feedback loops; Oscillators; Preamplifiers; Radiation detectors; Tuning; Varactors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1966.4324133
  • Filename
    4324133