DocumentCode
780637
Title
A rigorous treatment of self-shielding and covers in neutron spectra determinations
Author
Griffin, Patrick J. ; Kelly, John G Jake
Author_Institution
Sandia Nat. Labs., Albuquerque, NM, USA
Volume
42
Issue
6
fYear
1995
fDate
12/1/1995 12:00:00 AM
Firstpage
1878
Lastpage
1885
Abstract
Most neutron spectrum determination methodologies ignore self-shielding effects in dosimetry foils and treat covers with an exponential attenuation model. This work provides a quantitative analysis of the approximations in this approach. It also provides a methodology for improving the fidelity of the treatment of the dosimetry sensor response to a level consistent with the user´s spectrum characterization approach. A library of correction functions for the energy-dependent sensor response has been compiled that addresses dosimetry foils/configurations in use at the Sandia National Laboratories Radiation Metrology Laboratory
Keywords
dosimetry; foils; neutron detection; neutron spectrometers; shielding; correction functions; covers; dosimetry foil; exponential attenuation model; neutron spectrum determination; self-shielding; sensor response; Attenuation; Dosimetry; Gamma ray detectors; Gold; Laboratories; Libraries; Metrology; Neutrons; Sensor phenomena and characterization; Sensor systems;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.489230
Filename
489230