Title :
Cutting wax with an ultrasonically activated blade
Author :
Vaitekunas, Jeffrey J. ; Stulen, Foster B. ; Grood, Edward S.
Author_Institution :
Electr. Eng., Purdue Univ., Hammond, IN
Abstract :
In this paper, a model was developed for cutting through wax using an ultrasonic instrument in the shear mode. The model was based on viscous heating in a liquid layer of molten wax and extrusion of the molten wax away from an advancing blade, using an energy-balance approach. The model provides a set of equations representing cut rate as a function of velocity and normal force. Using two parameters in the model determined by a regression of experimental data at a single frequency, the model accurately reflected the behavior of experimental data at other frequencies and using blades of different geometries.
Keywords :
biomedical ultrasonics; extrusion; waxes; energy-balance approach; extrusion; instrument performance; shear mode; ultrasonically activated blade; viscous heating; wax cutting; Acoustic devices; Blades; End effectors; Heating; Piezoelectric transducers; Predictive models; Shape; Surgery; Surgical instruments; Ultrasonic transducers;
Journal_Title :
Engineering in Medicine and Biology Magazine, IEEE
DOI :
10.1109/MEMB.2008.924848