    Active D-C Restoration in Nuclear Pulse Spectrometry

  • Author

    Gere, E.A. ; Miller, G.L.

  • Author_Institution
    Bell Telephone Laboratories, Incorporated Murray Hill, New Jersey
  • Volume
    14
  • Issue
    1
  • fYear
    1967
  • Firstpage
    89
  • Lastpage
    95
  • Abstract
    It is well-known that while the use of single clipping yields optimum energy resolution in nuclear pulse spectrometry it suffers from the drawback of producing rate-dependent baseline shifts. On the other hand the use of double clipping removes the baseline shift but degrades the signal-to-noise ratio. A scheme has been devised which circumvents some of these problems by producing a signal proportional only to the peak-to-peak excursion of an input signal. When operating with double clipped signals this improves the signal-to-noise ratio and also reduces tail-pile-up difficulties. Single clipped signals can also be used and produce an output amplitude which is independent of baseline position, allowing the use of single clipping at high counting rates, although in this case the system exhibits a noise disadvantage. The technique employed is based on the principle of "active D-C restoration" and circuit details, together with representative results, are given.
  • Keywords
    Circuits; Diodes; Gamma ray detection; Gamma ray detectors; Germanium; Pulse amplifiers; Signal restoration; Spectroscopy; Switches; Telephony;