Title :
Reaction of truncation artifacts in fan beam transmission imaging using a spatially varying gamma prior
Author :
Case, James A. ; Pan, Tin-Su ; King, Michael A. ; Luo, Der-shan ; Penney, Bill C. ; Rabin, Monroe S Z
Author_Institution :
Massachusetts Univ. Med. Center, Worcester, MA, USA
fDate :
12/1/1995 12:00:00 AM
Abstract :
Using fan beam transmission imaging to obtain an attenuation estimate of an object for attenuation correction is limited because the object may not be completely in the field of view of the camera at all angles. This truncation of object data creates artifacts in the attenuation map during reconstruction which may result in problems in the attenuation correction of the emission image. To improve the quality of the attenuation map, the authors have used information from the scatter window reconstruction of projections from a parallel hole collimator on one of the heads of the SPECT system. This scatter window reconstruction is used to create a segmentation-based attenuation map of the object. The authors use this map as a pixel-by-pixel gamma prior to aid in the iterative reconstruction of the final attenuation map. In the fully sampled central region, use of the segmentation map information is unnecessary; but in the poorly sampled, outer region, the use of the segmentation map is essential to obtain an accurate reconstruction. The authors use a spatially varying weight function for the gamma prior to reflect their confidence in the transmission in the central region while emphasizing the prior in the outer region. The result is a high-quality density map over both the fully sampled and poorly sampled regions. Furthermore, since the line integrals for the transmission data are preserved in the iterative reconstruction, the algorithm does not require precise determinations of the attenuation coefficients in the segmentation maps to create an accurate attenuation map
Keywords :
image reconstruction; image segmentation; medical image processing; single photon emission computed tomography; SPECT system; accurate attenuation map; attenuation correction problems; camera field of view; emission image; fan beam transmission imaging; iterative reconstruction; line integrals; medical diagnostic imaging; nuclear medicine; object attenuation estimate; parallel hole collimator; scatter window reconstruction; segmentation-based attenuation map; spatially varying gamma prior; spatially varying weight function; truncation artifacts; Attenuation measurement; Cameras; Collimators; Electromagnetic scattering; Heart; Image reconstruction; Image segmentation; Optical imaging; Particle scattering; Single photon emission computed tomography;
Journal_Title :
Nuclear Science, IEEE Transactions on