DocumentCode
782899
Title
An analysis of correction methods for emission contamination in PET postinjection transmission measurement
Author
Luk, W.K. ; Digby, W.D. ; Jones, W.F. ; Casey, M.E.
Author_Institution
CTI PET Systems, Knixville, TN, USA
Volume
42
Issue
6
fYear
1995
fDate
12/1/1995 12:00:00 AM
Firstpage
2303
Lastpage
2308
Abstract
Three techniques for estimation of emission contamination in PET postinjection transmission scans (PITSs) were evaluated and compared. For PET scanners with electronic windowing of events that intersect the transmission rod source locations, the emission contamination can be estimated by: (1) the emission data taken either before or after the transmission scan, using a sinogram window mask to correct for the rod position and window width; (2) the emission data taken concurrently with the transmission data and a similar sinogram window mask; (3) utilizing a windowed transmission scan with cold rods, or “mock” scan. These techniques have been evaluated with phantom experiments to simulate a wide range of residual activities and object sizes
Keywords
positron emission tomography; PET postinjection transmission; PET scanners; cold rods; correction methods; electronic windowing; emission contamination; phantom experiments; positron emission tomography; postinjection transmission scans; residual activities; sinogram window mask; transmission rod source locations; window width; Attenuation; Contamination; Helium; Imaging phantoms; Oncology; Pollution measurement; Position measurement; Positron emission tomography; Time measurement; Whole-body PET;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.489432
Filename
489432