DocumentCode :
783327
Title :
Precision impedance measurements by the three-voltage method with a novel high-stability multiphase DDS generator
Author :
Callegaro, Luca ; Galzerano, Gianluca ; Svelto, Cesare
Author_Institution :
Inst. Elettrotecnico Nazionale Galileo Ferraris, Torino, Italy
Volume :
52
Issue :
4
fYear :
2003
Firstpage :
1195
Lastpage :
1199
Abstract :
The system for the intercomparison of unlike impedances of the Istituto Elettrotecnico Nazionale Galileo Ferraris, based on the three-voltmeter method and four-terminal pair definition of the standards, has now been provided with a new direct-digital synthesis generator, having multiple outputs, isolation among the channels, and high amplitude and frequency stability. Experimental results obtained using the new generator are shown and compared with those achieved with the previous system configuration.
Keywords :
direct digital synthesis; electric impedance measurement; measurement standards; voltmeters; amplitude stability; four-terminal pair standard; frequency stability; impedance measurement; multiphase direct digital synthesis generator; three-voltmeter method; Bridge circuits; Current measurement; DC generators; Electrical resistance measurement; Impedance measurement; Master-slave; Noise generators; Phase measurement; Stability; Voltage;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2003.815990