Title :
Spectral CT Modeling and Reconstruction With Hybrid Detectors in Dynamic-Threshold-Based Counting and Integrating Modes
Author :
Liang Li ; Zhiqiang Chen ; Wenxiang Cong ; Ge Wang
Author_Institution :
Dept. of Eng. Phys., Tsinghua Univ., Beijing, China
Abstract :
Spectral CT with photon counting detectors can significantly improve CT performance by reducing image noise and dose, increasing contrast resolution and material specificity, as well as enabling functional and molecular imaging with existing and emerging probes. However, the current photon counting detector architecture is difficult to balance the number of energy bins and the statistical noise in each energy bin. Moreover, the hardware support for multi-energy bins demands a complex circuit which is expensive. In this paper, we promote a new scheme known as hybrid detectors that combine the dynamic-threshold-based counting and integrating modes. In this scheme, an energy threshold can be dynamically changed during a spectral CT scan, which can be considered as compressive sensing along the spectral dimension. By doing so, the number of energy bins can be retrospectively specified, even in a spatially varying fashion. To establish the feasibility and merits of such hybrid detectors, we develop a tensor-based PRISM algorithm to reconstruct a spectral CT image from dynamic dual-energy data, and perform experiments with simulated and real data, producing very promising results.
Keywords :
compressed sensing; computerised tomography; image denoising; image reconstruction; medical image processing; photon counting; tensors; CT performance; complex circuit; compressive sensing; contrast resolution; dynamic dual-energy data; dynamic-threshold-based counting; energy bin; functional imaging; hardware support; hybrid detectors; image noise; integrating modes; material specificity; molecular imaging; multienergy bins; photon counting detectors; spatially varying fashion; spectral CT modeling; spectral CT reconstruction; spectral CT scan; spectral dimension; statistical noise; tensor-based PRISM algorithm; Attenuation; Computed tomography; Detectors; Energy resolution; Image reconstruction; Photonics; Tensile stress; Dynamic threshold; hybrid detector; image reconstruction; spectral CT; tensor-based PRISM;
Journal_Title :
Medical Imaging, IEEE Transactions on
DOI :
10.1109/TMI.2014.2359241