• DocumentCode
    786146
  • Title

    Flexible quadrature coil system for simultaneous reception of magnetic resonance signals [biomedical NMR]

  • Author

    Molyneaux, David A. ; Qureshi, Ahmed Hussain

  • Author_Institution
    NMR Sci. & Technol. Centre, Picker Int., Cleveland, OH, USA
  • Volume
    31
  • Issue
    6
  • fYear
    1995
  • fDate
    11/1/1995 12:00:00 AM
  • Firstpage
    3590
  • Lastpage
    3592
  • Abstract
    The magnetic field intensity of a flexible quadrature coil system for magnetic resonance imaging is modeled and simulated. The calculated signal to noise ratio of the coil system is verified experimentally using a coil constructed with the model´s dimensions. It is found that flexing the coil over a thirty degree range preserves the high signal to noise ratio and increases the clinical utility of such a system
  • Keywords
    biomedical NMR; biomedical equipment; coils; magnetic fields; magnetic noise; NMR imaging; clinical utility; coil flexing; flexible quadrature coil system; magnetic field intensity; magnetic resonance imaging; magnetic resonance signals; medical diagnostic imaging; signal to noise ratio; simultaneous reception; Coils; Humans; Magnetic fields; Magnetic resonance; Nuclear magnetic resonance; Phased arrays; Polarization; Radio frequency; Signal to noise ratio; Spectral analysis;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/20.489579
  • Filename
    489579