DocumentCode
788241
Title
The Possibilities of Using CdTe as a Gamma Spectrometer
Author
Akutagawa, W. ; Zanio, K.
Author_Institution
Hughes Research Laboratories Malibu, California
Volume
15
Issue
3
fYear
1968
fDate
6/1/1968 12:00:00 AM
Firstpage
266
Lastpage
274
Abstract
Photopeaks have been resolved from monoenergetic gammas within the 122 to 840 keV range using 40 mm3 surface barrier CdTe detectors operated at room temperature and under less than full collection conditions. Presently available CdTe exhibits heavy charge trapping and the gamma spectra reflect the effect of trapping and detrapping under changes of temperature, bias, and clipping time.
Keywords
Conductivity; Crystalline materials; Crystals; Energy resolution; Gamma ray detection; Gamma ray detectors; Laboratories; Photonic band gap; Spectroscopy; Temperature;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1968.4324947
Filename
4324947
Link To Document