• DocumentCode
    788434
  • Title

    An Anticoincidence-Shielded Ge(Li) Gamma-Ray Spectrometer for High Sensitivity Counting

  • Author

    Cooper, J.A. ; Wogman, N.A. ; Perkins, R.W.

  • Author_Institution
    Environmental and Radiological Sciences Department Battelle Memorial Institute, Pacific Northwest Laboratory Richland, Washington
  • Volume
    15
  • Issue
    3
  • fYear
    1968
  • fDate
    6/1/1968 12:00:00 AM
  • Firstpage
    407
  • Lastpage
    412
  • Abstract
    An anticoincidence-shielded Ge(Li) gamma-ray spectrometer for high sensitivity counting is described. The system consists of a 20 cc coaxial diode centered inside a 26" diameter by 24" thick plastic phosphor. The source is placed inside the shield and both the coincidence and anticoincidence spectra are recorded simultaneously. The anticoincidence shield reduces the 137Cs Compton edge by a factor of 7.5 to provide a peak to Compton edge ratio of 150:1 without reducing the photopeak efficiency by more than 3%. This system also provides a two to fourfold lower natural radionuclide background than similar sized diodes.
  • Keywords
    Coaxial components; Detectors; Diodes; Interference; Laboratories; Neutrons; Phosphors; Plastics; Radioactive materials; Spectroscopy;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1968.4324965
  • Filename
    4324965