DocumentCode :
788902
Title :
On the Origin of Leakage Current in Ge(Li) Gamma-Ray Detectors
Author :
McIntyre, R.J.
Author_Institution :
RCA Victor Company Ltd., Montreal, Canada
Volume :
15
Issue :
5
fYear :
1968
Firstpage :
6
Lastpage :
8
Abstract :
Hole-electron pair generation in Ge(Li) ¿-ray detectors due to thermal radiation from the cryostat outer casing is suggested as an often overlooked source of detector leakage current. Thermal-radiation-induced currents of the order of 10-9A are estimated for a typical large coaxial detector at 77°K when the cryostat outer casing is at 300°K.
Keywords :
Absorption; Coaxial components; Gamma ray detection; Gamma ray detectors; Leak detection; Leakage current; Radiation detectors; Reflectivity; Surface waves; Temperature dependence;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1968.4325009
Filename :
4325009
Link To Document :