Title :
On the Origin of Leakage Current in Ge(Li) Gamma-Ray Detectors
Author_Institution :
RCA Victor Company Ltd., Montreal, Canada
Abstract :
Hole-electron pair generation in Ge(Li) ¿-ray detectors due to thermal radiation from the cryostat outer casing is suggested as an often overlooked source of detector leakage current. Thermal-radiation-induced currents of the order of 10-9A are estimated for a typical large coaxial detector at 77°K when the cryostat outer casing is at 300°K.
Keywords :
Absorption; Coaxial components; Gamma ray detection; Gamma ray detectors; Leak detection; Leakage current; Radiation detectors; Reflectivity; Surface waves; Temperature dependence;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1968.4325009