• DocumentCode
    788902
  • Title

    On the Origin of Leakage Current in Ge(Li) Gamma-Ray Detectors

  • Author

    McIntyre, R.J.

  • Author_Institution
    RCA Victor Company Ltd., Montreal, Canada
  • Volume
    15
  • Issue
    5
  • fYear
    1968
  • Firstpage
    6
  • Lastpage
    8
  • Abstract
    Hole-electron pair generation in Ge(Li) ¿-ray detectors due to thermal radiation from the cryostat outer casing is suggested as an often overlooked source of detector leakage current. Thermal-radiation-induced currents of the order of 10-9A are estimated for a typical large coaxial detector at 77°K when the cryostat outer casing is at 300°K.
  • Keywords
    Absorption; Coaxial components; Gamma ray detection; Gamma ray detectors; Leak detection; Leakage current; Radiation detectors; Reflectivity; Surface waves; Temperature dependence;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1968.4325009
  • Filename