DocumentCode :
78961
Title :
Frequency Calibration of Terahertz Time-Domain Spectrometer Using Air-Gap Etalon
Author :
Kinoshita, Moto ; Iida, Hiroyuki ; Shimada, Yusuke
Author_Institution :
Nat. Metrol. Inst. of Japan (NMIJ), Nat. Inst. of Adv. Ind. Sci. & Technol. (AIST), Tsukuba, Japan
Volume :
4
Issue :
6
fYear :
2014
fDate :
Nov. 2014
Firstpage :
756
Lastpage :
759
Abstract :
Although various terahertz (THz) technologies have been rapidly developed in recent years, their metrology standards are yet to be established. We have developed a simple calibration method for the frequency of a THz time-domain spectrometer (TDS) using an optically transparent THz air-gap etalon (OT-TAGE). The OT-TAGE can be used in both the near-infrared (NIR) and THz regions. Therefore, a frequency measurement using THz waves can be directly compared with one using NIR light via the OT-TAGE. In this study, the frequency of a commercial THz-TDS was calibrated by referring to the NIR wavelength was demonstrated. The obtained frequency errors were within ± 3 GHz, with relative uncertainties within 1%.
Keywords :
calibration; frequency measurement; measurement errors; measurement uncertainty; terahertz spectroscopy; THz regions; frequency calibration; frequency errors; frequency measurement; near-infrared regions; optically transparent THz air-gap etalon; terahertz time-domain spectrometer; uncertainties; Air gaps; Calibration; Frequency measurement; Interferometers; Optical variables measurement; Time-domain analysis; Uncertainty; Time-domain analysis; air gaps; calibration; frequency; interferometers;
fLanguage :
English
Journal_Title :
Terahertz Science and Technology, IEEE Transactions on
Publisher :
ieee
ISSN :
2156-342X
Type :
jour
DOI :
10.1109/TTHZ.2014.2357895
Filename :
6905868
Link To Document :
بازگشت