DocumentCode :
790225
Title :
A Low Level Environmental Radiation Monitor with High Sensitivity, Stability, and Dynamic Range
Author :
Cassada, Thomas E.
Author_Institution :
Electronics Section Southwestern Radiological Health Laboratory
Volume :
16
Issue :
1
fYear :
1969
Firstpage :
405
Lastpage :
410
Abstract :
An environmental radiation monitor is described which records gamma radiation levels with a logarithmic response over a dynamic range of four decades (.01 mR/hr to 100 mR/hr achievable without range switching). Constant temperature accuracy is greater than ±10% of indicated reading throughout 3.5 decades of the four decade range. A high degree of temperature stability is achieved (error due to temperature change is less than ±20% of reading over an ambient temperature range of -10°C to + 55°C). Solid-state electronic circuitry is utilized exclusively for increased reliability and long term stability.
Keywords :
Detectors; Dynamic range; Filling; Gamma rays; Mercury (metals); Radiation monitoring; Solid state circuits; Stability; Temperature distribution; Voltage;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1969.4325136
Filename :
4325136
Link To Document :
بازگشت