Performance Comparisons of Continuous Miniature Crystal Element (cMiCE) Detectors
Author :
Ling, Tao ; Lee, Kisung ; Miyaoka, Robert S.
Author_Institution :
Dept. of Phys., Univ. of Washington, Seattle, WA
Volume :
53
Issue :
5
fYear :
2006
Firstpage :
2513
Lastpage :
2518
Abstract :
In this paper, we investigated the performance characteristics of continuous miniature crystal element (cMiCE) detectors. Versions with a 25 mm by 25 mm by 4 mm-thick LSO crystal and with a 50 mm by 50 mm by 8 mm-thick LYSO crystal were evaluated. Both detectors utilize a 64-channel flat panel photomultiplier tube (PMT). The intrinsic spatial resolution for the detectors was evaluated using Anger (i.e., simple centroid) positioning and a statistics based positioning (SBP) algorithm. We also compared the intrinsic spatial resolution for the 8-mm-thick LYSO crystal using different reflective materials (e.g., TFE Teflon, white paint, and a polymer mirror film) applied on the entrance surface of the crystal. The average energy resolution was 20% for the 4-mm-thick LSO crystal and ranged from 16% to 21%, depending upon reflective material, for the 8-mm-thick LYSO crystal. The average intrinsic spatial resolution for the 4-mm-thick crystal was 1.8-mm full width at half maximum (FWHM) for Anger positioning to within 3 mm of the crystal´s edge and 1.14-mm FWHM for SBP to within 2 mm of the edge. The average intrinsic spatial resolution for the 8-mm-thick crystal was 2.2-mm FWHM for Anger positioning to within 8 mm of the crystal´s edge and 1.3- to 1.5-mm FWHM (depending on reflective material used) for SBP to within 2 mm of the edge. Intrinsic spatial resolution is reported without correcting for point source size. The point spot flux had a FWHM of about 0.52 mm. The SBP algorithm resulted in significant improvement in intrinsic spatial resolution, linearity of positioning result, and effective field of view (FOV) for our cMiCE detector
Keywords :
photomultipliers; position sensitive particle detectors; positron emission tomography; solid scintillation detectors; 2 mm; 25 mm; 3 mm; 4 mm; 50 mm; 64-channel flat panel photomultiplier tube; 8 mm; Anger positioning; FOV; FWHM; LSO crystal; LYSO crystal; SBP; TFE Teflon; cMiCE; continuous miniature crystal element detectors; effective field of view; energy resolution; intrinsic spatial resolution; point source size; point spot flux; polymer mirror film; position sensitive detectors; reflective materials; simple centroid; small animal PET; statistics based positioning algorithm; white paint; Crystalline materials; Detectors; Energy resolution; Mirrors; Optical films; Paints; Photomultipliers; Polymer films; Spatial resolution; Statistics; Continuous crystal; position sensitive photomultiplier tube (PS-PMT); positioning algorithm; small animal PET;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2006.882296
Filename :
1710231
Link To Document :
بازگشت