  • Title

    Variation of the Effective Fano Factor in a Ge(Li) Detector

  • Author

    Sher, A.H. ; Keery, W.J.

  • Author_Institution
    Institute for Applied Techology National Bureau of Standards Washington, D. C. 20234
  • Volume
    17
  • Issue
    1
  • fYear
    1970
  • Firstpage
    39
  • Lastpage
    43
  • Abstract
    Variations in the value of the effective Fano factor, F, have been observed in Ge(Li) detectors using a collimated beam of ??-rays to irradiate selected areas in the sensitive region between the n+-and p-contacts. For regions near the n+-contact where the trapping of electrons is minimized, F was found to be statistically less than 0.11.
  • Keywords
    Charge measurement; Current measurement; Electric variables measurement; Energy measurement; Loss measurement; NIST; Noise measurement; Production; Pulse measurements; Radiation detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1970.4325556
  • Filename
    4325556