• DocumentCode
    796170
  • Title

    Soft X-Ray Studies of Channel Multipliers and Multiplier Arrays

  • Author

    Parkes, W. ; Gott, R. ; Pounds, K.A.

  • Author_Institution
    Department of Physics, University of Leicester, Leicester, England
  • Volume
    17
  • Issue
    3
  • fYear
    1970
  • fDate
    6/1/1970 12:00:00 AM
  • Firstpage
    360
  • Lastpage
    366
  • Abstract
    Measurements of pulse height distributions, gain and photon detection efficiency have been made for channel electron multipliers in the soft x-ray region. In the case of channel arrays, double exponential pulse height distributions have been found, and contact noise problems encountered. The variation of detection efficiency with angle of the incident beam and the uniformity of gain and efficiency over the cathode face have been studied. Results are given of a preliminary investigation of array operating lifetime. The detection efficiency over the cathode of a coned single channel was measured, and a comparison for angular variation between coned and plain cathode detectors is presented.
  • Keywords
    Anodes; Cathodes; Contacts; Electrons; Face detection; Pulse measurements; Radiation detectors; Spectroscopy; X-ray detection; X-ray detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1970.4325713
  • Filename
    4325713