DocumentCode
797615
Title
Slow Pulses from Germanium Detectors
Author
Sakai, Eiji
Author_Institution
Physics Division Chalk River Nuclear Laboratories Atomic Energy of Canada Limited Chalk River, Ontario, Canada
Volume
18
Issue
1
fYear
1971
Firstpage
208
Lastpage
218
Abstract
Slow pulses from various types of Ge(Li) detectors, i.e., a planar, a coaxial, a thin p+-contact planar, and a ´back-to-back´ double planar detector, have been studied by using a gamma-ray scanning technique and a risetime selector. A large and abrupt increase in the counting rates of the slow pulses were observed when the n+-i and the i-p boundaries of these detectors were irradiated by a collimated gamma-ray beam. Slow pulses contributed to background counts in the pulse height distributions. Slow pulses from a deeply depleted n+-p detector made from a high-purity p-type germanium crystal were also investigated.
Keywords
Charge carrier processes; Coaxial components; Electric potential; Gamma ray detection; Gamma ray detectors; Germanium; Laboratories; Physics; Rivers; Voltage;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1971.4325866
Filename
4325866
Link To Document