DocumentCode :
797702
Title :
A Laboratory Quality Charge Sensitive Amplifier Suitable for Space Applications
Author :
Alexander, R.M.
Author_Institution :
TRW Systems Redondo Beach, California
Volume :
18
Issue :
1
fYear :
1971
Firstpage :
264
Lastpage :
274
Abstract :
The amplifier described is designed for use with uncooled silicon nuclear particle detectors connonly used in space instruments. The charge sensitive preamplifier and two-stage shaping amplifier provides the performance of a laboratory quality instrument while simultaneously satisfying the stringent requirements imposed upon flight hardware. Double differentiation-double integration (DDDI) pulse shaping with pole-zero cancellation of the preamplifier decay time constant was selected on the basis of simplicity, adequate noise performance, and stabillity at high count rates. The components used in the design are compatible with hybrid microcircuit fabrication. Noise performance at 1 ??sec DDDI time constants was measured to be 1.8 KeV + 47 eV/pf FWHM (Silicon) for an amplifier consuming 100 mW. The operating temperature range is approximately -40?? to +80??C. Further results of an extensive testing program are reported.
Keywords :
Hardware; Instruments; Laboratories; Noise cancellation; Noise shaping; Preamplifiers; Pulse amplifiers; Radiation detectors; Silicon; Space charge;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1971.4325875
Filename :
4325875
Link To Document :