DocumentCode
803719
Title
Resolution uniformity and sensitivity of the NIH ATLAS small animal PET scanner: Comparison to simulated LSO scanners without depth-of-interaction capability
Author
Seidel, Jürgen ; Vaquero, Juan José ; Green, Michael V.
Author_Institution
Nat. Institutes of Health, Bethesda, MD, USA
Volume
50
Issue
5
fYear
2003
Firstpage
1347
Lastpage
1350
Abstract
Positron emission tomography (PET) scanners designed to image animals the size of rats and mice should possess simultaneously high and uniform spatial resolution and high sensitivity. ATLAS (Advanced Technology Laboratory Animal Scanner), a 6.0 cm diameter effective transverse field-of-view (FOV), 2 cm axial FOV ring-type research scanner seeks these goals by surrounding the animal with eighteen 15 mm deep, LGSO (7 mm)/GSO (8 mm) phoswich detector modules. A Monte Carlo simulation was used to compare the variation of resolution across the FOV and the absolute central point source sensitivity (ACS) of ATLAS to similar systems comprised only of LSO arrays of different depths with no depth-of-interaction (DOI) capability. For ATLAS radial spatial resolution deteriorated by 27% from the center to 3 cm off-axis. Scanners comprised of 15 mm deep, 10 mm deep and 7 mm deep LSO crystals deteriorated by 100%, 51%, and 20%, respectively, over the same distance. Simulated ACS (absorbed energies ≥250 keV) for ATLAS was 2.0% and for the 15 mm, 10 mm deep and 7 mm deep LSO scanners 2.4%, 1.5%, and 0.9%, respectively. Radial resolution loss 3 cm off-axis and ACS measured for the actual ATLAS scanner were similar to the values obtained by simulation (27% resolution loss, 1.8% ACS). The phoswich design thus achieves good resolution uniformity over a 6 cm FOV while preserving sensitivity compared to equivalent non-DOI LSO scanners with a range of crystal depths.
Keywords
Monte Carlo methods; biological techniques; positron emission tomography; 250 keV; ATLAS; GSO; LGSO; depth-of-interaction; field-of-view; phoswich; resolution uniformity; sensitivity; small animal PET scanner; spatial resolution; Animals; Crystals; Detectors; Laboratories; Loss measurement; Mice; Positron emission tomography; Pulse measurements; Rats; Spatial resolution;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2003.817282
Filename
1236930
Link To Document